7. Interferences: Acetaminophen, Dopamine,
L-DOPA, Ibuprofen, Salicylate, Bilirubin,
Creatinine, Cholesterol, Triglycerides, Uric
acid, EDTA, Galactose, Gentisic acid,
Glutathione, Haemoglobin, Heparin,
Maltose, Methyl-DOPA, Tolbutamide,
Tolazamide, Xylose, Pralidoxime Iodide and
Icodextrin (when occurring in normal or
therapeutic concentrations) do not
significantly affect results. However,
abnormal high concentrations of
substances, such as Acetaminophen,
Salicylate, Uric acid, and other reducing
substances in blood, may cause inaccurate
high results.
8. Intravenous administration of ascorbic acid
which results in blood concentration of
ascorbic acid >3 mg/dL will cause
overestimation of blood glucose results.
The Package of the GLM-77 Blood Glucose
Meter contains:
1. GLM-77 Blood Glucose Meter
2. Lancing Device with CE Marking
3. 3V Lithium Battery
4. User Manual
5. Carrying Case
NOTE: PLEASE MAKE SURE THAT ALL PROD
-
UCTS LISTED ON THE “CONTENTS” OF THE
BOX ARE CONTAINED IN THE PACKAGE
BEFORE USING THIS METER. IF YOU FIND ANY
IMPERFECTION IN OUR PRODUCTS, PLEASE
RETURN THE ENTIRE PACKAGE TO THE PLACE
OF PURCHASE.
4