<Toc> <2. Initial Settings>
2-5
IM 05F01D02-41E
Instrument Input Range Codes
Input Type
Instrument Input
Range Code
Instrument
Input Range
Measurement Accuracy
OFF
1
Set the data item PV Input Type "IN" to the OFF option to leave the PV input
type undefined.
2
K
J4
5
T
6
0.1% of instrument range 1 digit for temperatures
equal to or higher than 0°C
0.2% of instrument range 1 digit for temperatures
below 0°C
0.1% of instrument range 1 digit for temperatures
equal to or higher than 0°C
0.2% of instrument range 1 digit for temperatures
below 0°C
B7
0.15% of instrument range 1 digit for temperatures
equal to or higher than 400°C
5% of instrument range 1 digit for temperatures
below 400°C
S8
R9
0.15% of instrument range 1 digit
N10
0.1% of instrument range 1 digit
0.25% of instrument range 1 digit for temperatures
below 0°C
E11
L(DIN) 12
13
U(DIN)
14
W15
0.2% of instrument range 1 digit
Platinel 2 16
0.1% of instrument range 1 digit
PR20-40 17
0.5% of instrument range 1 digit for temperatures
equal to or higher than 800°C
No guarantee of accuracy for temperatures below 800°C
Thermocouple
Unspecified
W97Re3-
W75Re25
18
0.2% of instrument range 1 digit
30
0.1% of instrument range 1 digit (Note 1) (Note 2)
JPt100
31
0.2% of instrument range 1 digit (Note 1)
35
36
0.1% of instrument range 1 digit (Note 1) (Note 2)
RTD
Pt100
37
0.2% of instrument range 1 digit (Note 1)
0.4 to 2 V
40
Standard
signal
1 to 5 V 41
0 to 2 V 50
0 to 10 V 51
0.400 to 2.000 V
1.000 to 5.000 V
0.000 to 2.000 V
0.00 to 10.00 V
-10 to 20 mV
55 -10.00 to 20.00 mV
DC voltage
0 to 100 mV
56 0.0 to 100.0 mV
0.1% of instrument range 1 digit
The read-out range can be scaled between -1999 and
9999.
3
Set the data item PV Input Type "IN" to the OFF option to leave the PV input
type undefined.
-200 to 1370C
-300 to 2500F
-199.9 to 999.9C
0 to 2300F
-199.9 to 500.0C
-199.9 to 999.9F
-199.9 to 999.9C
-300 to 2300F
-199.9 to 400.0C
-300 to 750F
0.0 to 400.0C
-199.9 to 750.0F
0 to 1800C
32 to 3300F
0 to 1700C
32 to 3100F
0 to 1700C
32 to 3100F
-200 to 1300C
-300 to 2400F
-199.9 to 999.9C
-300 to 1800F
-199.9 to 900.0C
-300 to 1300F
-199.9 to 400.0C
-300 to 750F
0.0 to 400.0C
-199.9 to 750.0F
0 to 2300C
32 to 4200F
0 to 1390C
32 to 2500F
0 to 1900C
32 to 3400F
0 to 2000C
32 to 3600F
-199.9 to 500.0C
-199.9 to 999.9F
-150.0 to 150.0C
-199.9 to 300.0F
-199.9 to 850.0C
-300 to 1560F
-199.9 to 500.0C
-199.9 to 999.9F
-150.0 to 150.0C
-199.9 to 300.0F
* Performance in the standard condition (at 232C, 5510%RH, and 50/60 Hz power frequency.
Note 1: The accuracy is 0.3C of instrument range 1 digit for a temperature range from 0C to 100C.
Note 2: The accuracy is 0.5C of instrument range 1 digit for a temperature range from -100C to 200C.
* To receive a 4-20 mA DC signal, select a standard signal of 1 to 5 V DC and connect it to a 250 resistor. This
resistor is optional.
Model: X010-250-2 (resistor with M3.5 crimp-on terminal lugs)
1st Edition : May 31,2000-00