ZEUS-200S/150 ZO-OM-007
REV : 08 26 ZERONE
6. Periodic Inspection
Every six months, visually inspect the ZEUS-200S/150 Electrosurgical Generator for
signs of wear or damage.
In particular, look for any of the following problems:
Damage to the power cord
Damage to the power cable receptacle
Obvious damage to the unit
Damage to any receptacle
Accumulation of lint or debris in or around the unit
7. Technical Factors
7.1 Specification
Name of Product : Electrosurgical Unit
Model Name : ZEUS-200S / ZEUS-150
Rated Voltage : AC120V / AC230V
Rated Frequency : 50Hz / 60Hz
Power Consumption : 650VA +10% / 550VA +10%
Fuse : T6.3AL when AC120V / T4.0AL when AC230V
Class of Protection : Class 1, CF type
Neutral Electorde : Floating Insulated
Leakage current : in acc. with IEC601, Part2-2
Carrier Frequency : 400, 500
Repeat Frequency : 33
Size (Width × Depth × Height) : 330mm × 330mm × 135mm
Weight : 8.5Kg
Environment for storage, transport
Operation temperature : 10 to 40
Temperature : -10 to 60
Humidity : 20% to 95% RH,
Pressure : 700mbar ~ 1060mbar
Duty Cycle : 10sec ON 30sec Idle
Cooling : 1 inner fan
These parameters can be changed without notice.