EasyManuals Logo

Ziehm Imaging, Inc. VISION User Manual

Ziehm Imaging, Inc. VISION
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #29 background imageLoading...
Page #29 background image
Ziehm Imaging, Inc. Vision Installation Manual MAN 06-0015, Rev. A Page 16
Maintenance Report: Check Reproducibility
Fig. 6 Dosimeter Probe Placement
3. Select Radiographic Mode. Cover the intensifier input with 3 mm of lead blocker.
4. Set-up the x-ray system as follows:
kV = 60, mA = 15, ms = 200 for 120V operation.
kV = 60, mA = 20, ms = 200 for 230V operation
5. Take an exposure. Write the resulting dose value into the box below, on line one,
dose reading one, and reset the dose meter indicator to zero.
6. Repeat step 5 three times, writing the resulting dose values into line one of the
table below.
7. Change kV to 90.
8. Take an exposure. Write the resulting dose value into the box below, on line two,
dose reading one, and reset the dose meter indicator to zero.
9. Repeat step 8 three times, writing the resulting dose values into line two of the
table below.
* Note: If the system is wired for 230V
AC
, mA will be 20 not 15
Dose [mR]
Line kV mA ms 1 2 3 4 Avg.
1 60 *15 200
2 90 *15 200

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ziehm Imaging, Inc. VISION and is the answer not in the manual?

Ziehm Imaging, Inc. VISION Specifications

General IconGeneral
BrandZiehm Imaging, Inc.
ModelVISION
CategoryMedical Equipment
LanguageEnglish