9. APERTURE
28
4.7. Set position values according to the table below. Always navigate the aperture in positive direction to find positions. (False navigation
produces an error in the directions, and you will get false values!
4.8. By adjusting the Test count slide on Validation tab, you can define the number of tests to perform during validation. Click Start validation
test to validate the unit!
Type of objective
Aperture value
Pos. 1
20X
~0,5
Pos. 2
40X
~0,45
Pos. 3
40X wet
immersion
~0,45