8. Technical Maintenance
CAUTION
All technical service andrepair of theTRIOS IOS system
must be performed by3Shape or3Shape-authorized
personnelonly.
Theuser doesn’t need to perform any regular maintenance other than
calibration, cleaning, disinfectionandsterilization,asdescribed here.
Preventiveinspections andscheduled maintenance are not required.
System S1A(MOVE/MOVE+ cart) only
Changing thefuses (Figure 2, S20):
Shut down theMOVE/MOVE+ cart byfollowing theprocedure described
inSettingup & Starting TRIOS forthefirst time anddisconnect themains
power cable from theMOVE/MOVE+ cart.
CAUTION
Themains power cable must be disconnected before
changing fuses.
Using a pair of tweezers or similar, push in the two tabs on each side of the
fuse drawer simultaneously, as shown in the subfigure S20.
The drawer will begin to emerge from its slot.
It isthen possible toremove thedrawer andreplace thetwo fuses. Twospare
fuses are provided withthesystem ondelivery.
WARNING
Only fuses withthecorrect rating andtype
(250 V, T 3.15 AH) must be used.
Toreplace thefuses, first mount them inthedrawer, then insert thedrawer
back intotheslot andpush all theway inuntil thetabs click back intoplace
andthedrawer isno longer loose intheslot. Donotattempt toinsert
thefuses directly intotheslot ontheMOVE/MOVE+ cart.
Themains cable cannow be reconnected, andtheTRIOS IOS system startedup.
Changing thedongle for the MOVE/MOVE+ cart (software license
protection):
TheUSB wireless dongle, the USB license dongle and the USB wifi dongle,
and the Bluetooth dongle arelocated behind the plastic lid on top of the
backpack of the MOVE/MOVE+ cart. You canremove thelid byopening
thescrews holding thelid, asindicated bythearrows infigure S21.
WARNING
Thescanner must not be operated while theventilation
grid ontheback side of theMOVE/MOVE+ cart isremoved.
9. Consumable Accessories
Theconsumable accessories totheTRIOS IOS system are thescanner tips
andthecalibrationkit.
When thescanner tips are disposed, new tips canbe ordered bycontacting
your 3Shape reseller. Also, anew calibration
kit canbe ordered from your 3Shape reseller.
10. Vigilance
In case of any serious incident that has occurred in relation to the use of the
device, please report the incident to 3Shape at email: support@3shape.com
and you can also report it to the competent authority of the state in which
the user and/or patient is established.