K-METRON Radio User Instructions (Firmware 3.0.4; Doc v4.0)
10 | P a g e
4.3 Validating Calibration.
After calibration is complete the measured data should be validated. To do this position the
calibrated magnet once more on each step and ensure the K-METRON’s measured and
displayed value is that of the actual step thickness (note the measurement tolerances as per
the table above apply).
Magnet Used
Mag Type
Calibration Range
10mm Sphere
Sphere
0mm - 20mm
10mm Cube
Cube
0mm - 20mm
19mm Sphere
10-30mm
10mm 30mm
10mm Cube and 10mm Sphere Validating Calibration
19mm Sphere Validating Calibration use the provided +10mm disk to measure units
over 20mm.
+10mm Calibration
Add-on