OPERATION
General
This product is designed to provide maximum comfort to patients. Make sure that you operate this
product in a proper way optimizing its value. Here we provide some general information that you should
be aware of.
NOTE: Please follow instructions below for detailed operating procedure.
Step 1: Turn on the power. The indicator of the power key will flash.
Step 2: The pump will first default mode upon startup in the Alternating Pressure Mode to begin
providing alternating pressure to the ATLAS Mattress. Refer to the specific controls on page 5 & 6 of this
manual.
Step 3: Refer to the specific pump instructions (page 5 & 6) to enter the patient weight and desired
alternating cycle.
NOTE: Press the static mode button from the touch panel to provide a firm surface that makes it easier
for the patient to transfer or reposition. The static mode can support the patient by keeping the mattress
from bottoming out when the patient is in a sitting position. The static mode will revert to alternating
mode after 20 minutes. Press the Static button again to switch back to alternating mode.
Step 4: During normal operation, the unit will monitor pressure. If the mattress pressure is lower than the
set pressure, the pump will automatically inflate the mattress to the set level and then stop.
NOTE: In case there is a bad connection of tubing or any air leakage from the mattress, this will result in a
low-pressure condition, and it will activate the alarm within 1 minute. Press the Alarm Mute button to
mute the alarm. The alarm LED will continue flashing. Press the Alarm Mute button again to re-enable
the audible alarm.
CPR Function
CPR should be conducted to standard protocol for a standard hospital foam mattress.
Weight/Pressure Set Up
Users can adjust air flow to the mattress to a desired firmness according to the user’s weight or the
suggestion from a healthcare professional.
Low Pressure Warning
When abnormal pressure occurs, the Low-Pressure indicator will come on. For some types of pumps, the
alarm will be activated to alert of a low-pressure condition. Check if the connections are secure and
correctly installed at the pump and the mattress.
NOTE: If the pressure is consistently low, open the zipper and confirm that all hoses are properly
connected to the mattress and the pump. Then check for any noticeable leakage or holes in the tubing. If
necessary, contact your local dealer to replace any damaged tubes or hoses.