CLEANING
In this section, we describe the procedures to clean and decontaminate the pump. It is important to
follow these procedures before using the system again. The cleaning task is required at least once a
week to maintain personal hygiene.
Pump Unit and Hose Set
DO NOT immerse or soak the pump in any water or fluids.
Check for external damage and move the pump to the cleaning area.
Place the pump on a work surface and spray or wipe the outside of the case with quaternary
ammonium solution.
DO NOT spray any cleaning solution directly on the surface of the pump.
DO NOT use a Hypocarbonate or Phenolic based cleaning solution as this may cause damage to
the case.
Allow the solution to incubate for 10 minutes or accordingly as stated by the cleaning product
instructions.
Wipe the case with a clean cloth. Make sure all areas are clean (top and bottom, both sides).
Spray cloth with cleaning solution and clean faceplate. DO NOT allow excess cleaning solution on
the faceplate or control panel. (If solution gets inside, damage will occur.) Allow the surface to
thoroughly dry after cleaning.
After the pump is thoroughly cleaned and dried, proceed to plug in the pump and test to see if it
runs normally.
Unplug the pump and store with proper identification tag.
Avoid long exposure to sunlight.
Mattress
Washing and Disinfection General Guidance:
Because of the variety of laundry equipment, chemicals, and conditions in use, you should pre-
test cleaning of the cover with your agent of choice to ensure compatibility.
Routine Care: Surface wipe with neutral detergent and lukewarm water. Rinse thoroughly with
water and dry thoroughly.
Disinfection: Surface wipe the cover with quaternary solution or 1:10 bleach/water mixture
according to recommended protocols. Avoid over saturation and ensure the surface does not
stay wet longer than the chemical manufacturer’s guidelines for proper disinfecting. Wipe up
excess disinfectant after the recommended kill time has elapsed. Thoroughly rinse the surface of
the mattress with clean water using a clean cloth. Allow the surface to completely air dry prior
to use. Phenolics may be used for the vinyl bottom of the cover only
Storage: When not in use, mattress should be stored vertically in a controlled climate room.