What to do if A&D Software is showing Missing Files?
- Cchristopher18Sep 23, 2025
Check the folder specified in the Recorder-Retrieved Data dialog box is the same folder as specified using the File | Open function.
What to do if A&D Software is showing Missing Files?
Check the folder specified in the Recorder-Retrieved Data dialog box is the same folder as specified using the File | Open function.
What does E02 mean on A&D Doctor Pro TM-2430-13, and how to fix it?
Error E02 on A&D Software indicates an air leak. To resolve this, check that the air hose plug is tightly connected to the air connector at the front of the monitor. Also, check the connection between the air hose plugs and the air bladder, and inspect the air bladder for leaks.
What to do if A&D Software displays error E01?
If the A&D Software displays error code E01, it means that the microphone has become disconnected or there has been excessive body motion. The suggested solution is to remove and reconnect the microphone connector at the front of the monitor.
What does error E06 mean on A&D Doctor Pro TM-2430-13 Software, and how can I fix it?
Error E06 on A&D Software indicates that the pressurization is greater than 320 mmHg. To fix it, confirm that the subject is not moving during the measurement and clear the memory (note: all measurement data will be erased). After clearing the memory, take one measurement; if the cuff pressurizes to approximately 195 mmHg, the system is OK. Reset the monitor by gently inserting a paper clip in the hole located on the bottom of the monitor (note: all measurement data, as well as programmed intervals and conditions will be erased). If the monitor stops and then continues to inflate above 260 mmHg, return the unit to A&D for service.
What to do if A&D Doctor Pro TM-2430-13 Software Summary Data / Sleep Tab Data can not be Seen?
Make sure that the Sleep parameters in the Patient Information screen are within the monitoring session’s time.
What to do if A&D Doctor Pro TM-2430-13 Software shows error E32?
If A&D Software shows error E32, which indicates a clock error, repair is necessary if the error persists.
What to do if A&D Software shows error E75?
If A&D Software shows error E75, which indicates a protocol error due to external equipment, re-connect the communication cable. If the error persists, repair is necessary.
What to do if A&D Software shows error E30 or E31?
If A&D Software shows error E30 or E31, which indicates air leaks in the cuff assembly, repair is necessary.
What does error E03 mean on A&D Doctor Pro TM-2430-13 Software?
If A&D Software displays error E03, which means an error code is displayed without cuff inflation, exhaust the air from the cuff completely.
What to do if A&D Doctor Pro TM-2430-13 Software shows error E74?
If the A&D Software shows error E74, which indicates that voltage decreases during communication, replace the batteries with new ones and restart communication.
Overview of the ABPM Data Analysis Software, its features, and supported devices.
Details computer specifications and provides step-by-step instructions for installing the software.
Outlines essential safety precautions to be observed when connecting the monitor and processor to a computer.
Guides on starting the software and establishing communication between monitors and the computer.
Describes the main screen layout, menus, and toolbar buttons for software operation.
Outlines functions for managing data files: Open, Save As, Convert, Merge, and Print.
Explains how to access and view data through Summary Data, BP Data List, and Trends Screens.
Details Correlation Plots, Histograms, Circadian Rhythm, and HIB Analysis for data interpretation.
Covers managing patient information, retrieving data, and clearing monitor memory.
Explains how to set monitor intervals, conditions, and examples for different models.
Guides on creating reports, managing physician lists, and setting default software options.
Lists error codes for TM-2420/TM-2421 and TM-2430 series, with causes and corrections.
Describes steps to resolve communication errors between the monitor and PC.
Provides checks for missing files and issues with report printing based on BP Mode.
Shows sample printouts for patient info, summary data, trends, correlations, and histograms.
Defines terms and abbreviations related to blood pressure measurement, like Oscillometric and Korotkoff.
Defines statistical terms and abbreviations used in data analysis, such as Mean, SD, and CV.
Defines terms related to biological rhythm analysis, including Cosinor, MESOR, and Acrophase.
Defines terms like HBI, Hyperbalic Index, and Hypobalic Area for managing high blood pressure risk.