What to do if A&D Software displays error E01?
- MMelanie AllenAug 26, 2025
If the A&D Software displays error code E01, it means that the microphone has become disconnected or there has been excessive body motion. The suggested solution is to remove and reconnect the microphone connector at the front of the monitor.