Deep Cleaning Cleaning & Maintenance Procedures
2. Quality control.
Go to Maintenance> Quality Control> QC Normal. Make sure the control has come to
room temperature. Mix well by rolling the tube flat between your hands 10 times (A)
and then slowly invert the tube 10-15 times (B). Run the well mixed HM5 control 2-3
times and ensure the results are within range to confirm the analyzer is ready to be
used.
2a. If 2 out of 3 QC runs fail, proceed to Calibration.
Go to Maintenance> Calibration> Measure> Accept, then enter in the values
from the Package insert. Accept> Run Cal. After a successful calibration is
completed, go to Maintenance> Quality Control> QC Normal and run QC again.
Note: If a warning message pops up or the Quality control fails, call Abaxis Technical
Support.
A
B
10x
10x
5X