EasyManua.ls Logo

Abaxis HM5 - Warning Flags

Abaxis HM5
18 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Blank Warning Flags
Flag Meaning Solution
p PLT blank is high Visually inspect the wash head and clean if necessary; run 2-3
automatic self cleanings, then repeat the blank. If the warning
persists, run Soak Cleaning and the repeat the Blank.
X EOS blank high Visually inspect the wash head and clean if necessary: check
Diluent (green) and Lyse2 (orange) lines for loose connections;
prime Diluent and/or Lyse2 if bubbles are present or were
introduced while checking the tubing; repeat the blank.
b RBC blank is high Visually inspect the wash head and chamber below it for
blood debris; clean if necessary. Run a Soak Cleaning, then an
automatic self-cleaning; repeat the Blank.
More thorough troubleshooting procedures can be found in the Abaxis HM5 Operator‘s Manual.