7. From the New QC Lot screen, select the method
for entering the control target and ranges:
Enter Manually allows the user to type
in the target and range values from the
control package insert sheet.
Load from USB Drive allows the user to
upload the values downloaded onto a
USB drive from http://www.abaxis.com/
veterinary/products/vetscan-hm5
Scan Barcode allows the user to scan
the barcode on the control package
insert sheet using a 2D barcode scanner
to upload the values.
8. Accept the new values.
9. Confirm that the values entered are correct.
10. Follow the mixing instructions, remove
cap, and place tube in the sample adapter.
Touch Run.
11. Run the control 2-3 times to make sure
values are similar from run to run. If all
values reported are within reference range,
the analyzer is in calibration and samples
may be run. If values are repeatedly out of
range, consult Abaxis Technical Support
before calibrating.
(See Calibrating the Analyzer)