5 - Receive Setup: Pressing
puts the handheld in setup mode, so that the handheld can connect to the
i-STAT 1
Wireless Setup Utility.
After pressing , the user is asked for a password. After entering the appropriate password or pressing
E
NT, the user will see a Waiting for Setup screen. Once the test is complete, a Setup Successful or
Setup Failed message (with an error code) will appear.
Enter Password
_ _ _ _ _
W
aiting For
Setup
. . . .
1 – Cancel
Setup
S
uccessful
1 – Continue
Note: Pressing 1-Continue after
the Receive Setup
screen
navigates the user back to
the Wireless Menu.
5
5
Note 5.1: The Receive Setup function should only be used by IT personnel using the instructions
found in the Configuring Wireless Settings in an i-STAT
®
1 Wireless Analyzer Technical
Bulletin (Art: 726066).
i-STAT is a registered trademark of the Abbott Group of Companies in various jurisdictions.
Art: 726025-01A12 Rev. Date: 24-NOV-14
Wireless Module
Error Code Description
FCC ID: YOPGS1500M
Wireless module is not configured GS000
DHCP Process Time Out GS001
Connection to Access Point has failed GS002, GS010
Failed to make a TCP connection to Data Manager Port 6004 GS003
Unable to receive setup from APOC Wireless utility ad hoc GS020, GS021
Wireless configuration storage failed to erase/ write to flash GS022,GS023
Wireless configuration file corrupt or empty GS024, GS025
Device customization failed to enable wireless module GS030
Failed Ping request TIMEOUT
ERROR CODE MESSAGES

Other manuals for Abbott i-STAT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT and is the answer not in the manual?

Abbott i-STAT Specifications

General IconGeneral
TypeHandheld blood analyzer
Power SourceRechargeable lithium-ion battery
DisplayTouchscreen LCD
ConnectivityWireless
Sample TypeWhole blood
Test Cartridge TypesCG4+, CHEM8+, EG7+
Operating Temperature15° to 30°C (59° to 86°F)
Battery LifeApproximately 8 hours
Humidity Range10% to 90% non-condensing
Sample Volume95 µL
Time to Results2 minutes
Regulatory ApprovalsFDA cleared, CE marked
Measurement ParametersBlood gases, electrolytes, chemistries, coagulation, hematology

Related product manuals