A. Customizing the Handheld to Enable the Wireless Functionality Using the Handheld Keypad
1. Press to turn on the handheld.
2. Press
t
o change screen to Administration Menu.
3. Press (Customization).
4. Press (Change).
5. Enter password and press , or just press
if no password is required.
6. Press (Analyzer).
7. Press twice.
8. Press (Wireless).
9. Press to change setting to Enabled.
10. The “Wireless” listing should now appear as Enabled.
11. Press to turn handheld off and save the settings.
(If setting
is not shown as Enabled, return to Step 9).
B. Customizing the Handheld to Enable Wireless Functionality Using CDS Version 5
1. Click on Main ➔ Open Administration Function ➔ Customization.
2. Type in your password and click OK. The default password is the word istat.
Note: Abbott Point of Care Inc. recommends changing the default password.
M
ENU
4
1
➔
1
ENT
3
ENT
3
1
5:26 18JAN10
Administration Menu
1 - Analyzer Status
2
- Data Review
3 - Quality Tests
4 - Customization
5 - Set Clock
6 - Transmit Data
7 - Utility
Customization
Default0
1 – View
2 – Change
2
2
Art: 726025-01ARev. Date 24-NOV-14 3
Main System Tools Window Help
Instrument/Location
Operator
Database Maintenance
Inventory
Customization
User Administration
Open Monitor
Open Administration Function
Open Data Viewer
Open Report
Close
Exit Alt+X
X
X
X
X
Password System Tools Window
X
A password is needed to access this workspace.
Password:
OK Cancel
Change
Customization
Default0
1 – Analyzer
2 – ID Entry
3 – Patient Tests
4 – QC Tests
5 – Results
6 – Password
7 – Restore Factory
Settings
A
nalyzer
Customization
1
Wireless
Disabled
Change
Customization
Wireless
1 – Disabled
2 – Enabled