i1000SR System Quick Troubleshooting Guide
Error Code 3700 Unable to Process Test, (x)
Wash Aspiration Error for Probe(s) (y)
PN 205854-101
©2009, Abbott Laboratories, Abbott Park, IL, 60064
Refer to the Operations Manual for further instructions
Customer Support: 1-877- 4ABBOTT (1-877- 422-2688)
Instrument SN: ______________
Remove fibrin clots from sample cup or tube with a
clean applicator stick.
Perform the appropriate as-needed maintenance
procedure:
9 6445 Pipettor/WZ Probe Cleaning
Replace the wash zone probe.
Replace the wash zone temperature tubing and
sensor.
Wash Zone Tubing and Probe Assembly
Touchscreen Monitor fails to respond
Cycle power to System Control Center (SCC) and to the
Processing module by performing the following:
Select F3-Shutdown on the Snapshot screen. A
confirmation message displays.
Select OK to confirm the shutdown.
Wait for the information window to display, and then
simultaneously press the CTRL+ALT+DELETE keys
on the keyboard. The Confirm Exit window displays.
Leave the Shutdown Computer option selected,
select OK, and then wait for the information window
to display.
Turn off the power to the processing module power
supply by moving the power switch down.
NOTE: The Module Power
Switch is located on the
back side of the Processing
Module.
If touch-screen connector is loose or disconnected,
reseat or reconnect the touch-screen connector.
[1] monitor power cable
[2] touch-screen cable
[3] monitor video cable
Press the power switch on the front of the CPU to
turn on the SCC.
Wait for the Snapshot screen to display. It may
take several minutes to display.
Ensure the processing module(s) has been
powered off for one minute, then move the module
power supply power switch up to turn on power.
123
Page 1 of 2
Locate the Central Processing
Unit (CPU).
NOTE: Open the card cage and
SCC center door to access the
power switch.
Press and hold the power switch
on the front of the CPU to turn off
power to the SCC.
NOTE: The SCC may power off
immediately, or it may take up to
10 seconds depending on the
type of SCC you have.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i1000SR and is the answer not in the manual?

Abbott i1000SR Specifications

General IconGeneral
TypeImmunoassay Analyzer
ThroughputUp to 100 tests per hour
Assay TechnologyChemiluminescent Microparticle Immunoassay (CMIA)
Sample TypesSerum, plasma
Reagent CapacityUp to 25 reagents
CalibrationAutomatic
Quality ControlOnboard quality control
Assay MenuInfectious disease, cardiac, thyroid, fertility
Power Requirements100-240 VAC, 50/60 Hz