Language | English |
Mount unit vertically with brass fittings pointing down, using two #8 x 3/4" screws for optimal condensation management.
Horizontal mounting may cause zero shift. Prevent moisture buildup. Limit tubing to 75 feet for best results.
Use shielded 16-22AWG cable. Insert wires into depluggable terminal blocks, tightening screws.
Details on supported supply voltages (16-36 VDC, 24VAC) and output signal types (4-20mA, 0-5 VDC, 0-10 VDC).
Covers temperature ranges (0-85°C), humidity (10-95% RH), thermal effects, warm-up, and response times.
Specifies proof/burst pressures for different ranges and compatibility with dry air or inert non-conductive gases.
Includes DIN rail mounting, wire size (14-24 AWG), terminal torque, tubing size, and enclosure material (Polycarbonate).
ACI MLP2 sensors are covered by a Five (5) Year Limited Warranty.
Guidance on proper disposal of packaging and product at end-of-life to prevent environmental harm.