OMNISTIM
®
500 PRO
Electrotherapy System
User Manual
Revised 01.22.2018
Part No. 290500C Rev 5
This manual and all content therein is owned exclusively by Accelerated Care Plus ("ACP") and is protected by copyright. This manual or any
portion thereof may not be photocopied, reproduced or translated to another language without the express prior written consent of ACP. This
manual may only be used by entities that have purchased the equipment or have implemented the ACP program and are covered by an executed
lease agreement. This manual may not be used for any other purpose.
Any additional copies of the Manual shall be ordered from ACP. No changes or modifications shall be made to the Manual without prior review
and written authorization from ACP. No authorization is given to market, sell, disclose, or exploit this Manual except as for purposes of using the
Equipment as contemplated by the Lease Agreement.
ACCELERATED CARE PLUS MAKES NO WARRANTY OF ANY KIND WITH REGARD TO THIS MANUAL, INCLUDING, BUT NOT
LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Accelerated
Care Plus shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance
or use of this Manual.
The information contained in this document is subject to change without notice.

Questions and Answers

  • J
    Julie JonesAug 7, 2025
    What to do if my ACP OMNISTIM 500 PRO patient feels surging sensation?
    • A
      Andrew KellyAug 7, 2025
      If a patient using ACP Medical Equipment experiences a surging or spiking sensation, it could be due to several reasons. First, check the conductive medium interface; it might be absent, inadequate, or improperly applied, so replace it with the correct and adequate conductive medium. Another potential cause is a short or breakage in the lead wires; use a LEAD WIRE TESTER to check the wires and replace them if faulty. Also, the electrodes might be non-conductive or poorly conductive; remove the electrode(s) and replace if necessary.
  • R
    Rachel OwensAug 11, 2025
    What to do if my ACP OMNISTIM 500 PRO Medical Equipment unit will not power on with line power?
    • P
      Phillip ColeAug 12, 2025
      If your ACP Medical Equipment unit isn't powering on when using line power, first, make sure the power supply is properly plugged into both the unit and the AC outlet. If it is, verify that the AC outlet is functional and that the power plug being used is appropriate and undamaged. Finally, inspect the power supply operation.
  • T
    timothy13Aug 15, 2025
    What to do if my ACP Medical Equipment patient cannot detect output?
    • K
      Karen BrownAug 15, 2025
      If a patient using ACP Medical Equipment cannot detect output, it might be due to a failure of the lead wire(s), electrode(s), or conductive medium interface. Use an Omnistim® Output Tester to determine if the unit has failed or is operating incorrectly. If the LED illuminates properly, test the lead wire(s). If the LED is not illuminated, this indicates a failure of the Omnistim® 500 Pro. If the Output Tester shows the unit to be functional, and the Lead Wire tester show the lead wires to be functional, examine the electrodes, or conductive medium interface for problems.
  • E
    Elizabeth WalkerAug 20, 2025
    How to fix ACP OMNISTIM 500 PRO Medical Equipment that will not power on with batteries?
    • C
      Caitlin PooleAug 20, 2025
      If your ACP Medical Equipment unit doesn't power on when using batteries, the problem could be due to a few reasons. First, ensure that batteries are installed. If batteries are installed, they might be low; charge or replace the batteries. Verify the type and settings of batteries and inspect the battery contacts and setting.
  • L
    Luke FosterAug 25, 2025
    Why does my ACP OMNISTIM 500 PRO Medical Equipment display show a low battery warning?
    • G
      gphillipsAug 25, 2025
      If your ACP Medical Equipment displays a “LOW BATTERY” warning, it means the battery charge is too low. Recharge the batteries for future use.
  • T
    toni32Aug 28, 2025
    What to do if ACP OMNISTIM 500 PRO unit will not start?
    • S
      Samantha FloresAug 28, 2025
      If your ACP Medical Equipment unit won't start, it might be because no program is selected. Select PROTOCOL to start the unit.
  • C
    Cory Miller DDSSep 2, 2025
    What to do if I cannot set carrier frequency on my ACP Medical Equipment?
    • K
      Kevin VillanuevaSep 2, 2025
      If you cannot set the carrier frequency on your ACP Medical Equipment, it is because a treatment is in progress. Stop the current treatment, make your adjustments, and then restart.

ACP OMNISTIM 500 PRO Specifications

General IconGeneral
Channels4
Therapy ModesTENS, EMS, Interferential
WaveformInterferential
Power SourceRechargeable Battery
Frequency Range1-150 Hz
Intensity Levels0-100 mA