The Belt Loop Plate may be used to guide the belt when the device is strapped to a fixed or
immobile object.
Cleaning and Disinfecting the Device
To prevent the spread of infection and diseases, the Activforce 2 device and pad should be
disinfected between each use. The Activforce 2 device external surfaces, including the rounded
pad and the soft, arched pad silicon sleeve, can be hand wiped at room temperature with a soft
cloth and a cleaning solution consisting of 70% isopropyl alcohol and water. After application,
the cleaning solution should be allowed to evaporate completely.
Alternatively, a cleaning solution of 1% hypochlorite and water (10,000 ppm) can also be used
to wipe the device, pads and silicon sleeves. Antibacterial hand sanitizing agents such as non-
alcohol wipes or alcohol based hand gels may also be used. Any residue from these agents
should be wiped away using a soft, dry clean tissue or towel.
Replacing the Battery
The Activforce 2 device comes with a standard AAA battery, which lasts between six to twelve
months, depending on usage and environmental conditions.
To replace the battery, follow these steps:
Obtain a new AAA battery and a medium size flat-head screwdriver