Replacing Faceplates (618 Models)
Removal: Hold the chestpiece with one hand.
Pop off the old plate diagonally from behind.
Replace: Position the bottom of the new plate to the socket of the
chestpiece. Snap the top firmly into place.
Replacing Eartips
To replace Adsoft
TM
Plus Eartips: Slide Adsoft
TM
PLUS eartip over adapter until it
locks tight into internal channel. Repeat steps for second aural tube.
Cleaning, Disinfection, and Storage
Clean your Adscope between patients using either a 70% isopropyl alcohol
solution, 70% isopropyl alcohol wipe, or disposable wipe with soap and water.
Be sure to wipe down all surfaces including chestpiece, tubing, headset and eartips.
The diaphragm, non-chill bell ring, and eartips may be removed for a thorough
cleaning if desired. Ensure all parts are thoroughly dry before reassembling.
Do not remove chestpiece.
If storing in original box, the molded tray (all models except 609) may also be
wiped down to reduce risk of re-contamination.
Avoid extreme heat, cold, solvents or oils.
Do not leave exposed to direct sunlight for prolonged periods of time.
Periodically check that eartips are securely affixed to the headset and freefrom
dam-age.
Do not immerse stethoscope in any liquid.
Do not steam sterilize. If sterilization is necessary gas sterilize only.