Questions and Answers

  • A
    Amy BoothAug 12, 2025
    What to do if the ADC Diagnostix E-sphyg 7002 cuff inflates to over 300 mmHg?
    • C
      changinaAug 12, 2025
      If the ADC Blood Pressure Monitor cuff inflates to over 300 mmHg, immediately open the valve to rapidly exhaust the air from the cuff.
  • D
    Dillon BuckAug 18, 2025
    Why does my ADC Blood Pressure Monitor show a heartbeat rate detection error?
    • T
      Tracy StuartAug 18, 2025
      If you see a heartbeat rate detection error on your ADC Blood Pressure Monitor, it might be because the cuff deflated too rapidly. Ensure the deflation rate is approximately 2-3 mmHg per second.
  • J
    Jerry AdkinsAug 27, 2025
    What to do if the ADC Diagnostix E-sphyg 7002 Blood Pressure Monitor detects a pressure of over 15mmHg remaining for more than 3 minutes?
    • D
      dillonjonesAug 27, 2025
      If the ADC Blood Pressure Monitor displays a pressure of over 15mmHg for more than 3 minutes, press the power button to switch the unit off. The unit will automatically switch off after 60 seconds.
  • R
    Robert Mullins DVMSep 3, 2025
    How to fix a data error on an ADC Diagnostix E-sphyg 7002?
    • A
      alexis85Sep 4, 2025
      If the ADC Blood Pressure Monitor shows a data error, remove and reinsert the batteries.