7.3 Calibration Process
For a calibration example we will use a (0~100) kPa gauge.
(1) Access the calibration menu in Menu setup mode(refer to Menus operation section),
Figure 1 will display on the screen. Next push button to enter Menu setup,
Figure 2 will be displayed. Input password218(refer to Number inputting introduction
section) as in Figure 3,and push button to continue, Figure 4 will be shown.
(2) Select and push the screen will display the lower range of this gauge as
shown in Figure 5. The first digit will blink which allows the user to configure the
calibration point by inputting the desired pressure value. If no adjustment is needed,
just push to continue;
zero
zero
zero
zero