Custom Unit Entry
9
13
(1)
(2) Press key until you see TOOL at the bottom. Then press to enter the TOOL menu.
(3)
(Number x kPa). The default number is 1.
The flashing number is what can be adjusted. Note that at this point we are not modifying the
decimal, just the main integers (i.e. ones, tens, etc). The decimal adjustment will be step 6.
(5) Use key to scroll to the desired number. Then use the BACK LIGHT key to
sequence the number up. When you have the desired number press to enter the
value.
(6) Now you will see 0.0000 with the first decimal flashing. Adjust the number in the same way as
step 5 and press to accept the value.
(7) Now you should see the adjusted value on the display. This is the custom unit factor.
Important Note: If the value returns to the default number of 1, then it is likely the value you
tried to enter is not supported by the 5-digit resolution of the gauge. For example, the