8.4 Faults and error messages
8.4.1 Faults
Fault Cause Measure
The display shows nothing. The scale has switched off
automatically.
"Tap" or switch on the scale.
The display shows nothing. The main switch is not switched
on.
Switch on the main switch.
The display shows nothing. Not connected to the mains. Use the supplied mains adapter
to connect the scale to the
mains.
The display shows nothing. The battery is empty. Insert new batteries.
The display shows nothing. No battery is inserted. Insert batteries.
The display indication
shows cryptic characters.
The display is in "Turning mode". Press the Turning key.
The scale wobbles. The scale is not levelled properly. Adjust the support feet. Use the
spirit level to check the
horizontal position.
8.4.2 Error messages
Error message Description Troubleshooting
[ Lo ] Empty battery. Insert new batteries.
Use the scale in mains
operation.
[uLoad] Underload (-20d)
Switch the scale off and on
again.
[oLoad] Overload (-9d)
Unload the scale. Zero range or
weighing capacity exceeded
[no 0.00] No zero point available. Zero the scale.
8.5 Maintenance
To ensure correct measurement, maintenance and repair should only be carried out
by authorised personnel.
To prevent the intended level of accuracy the product must be set up carefully and services
regularly. We recommend having it serviced every 3 to 5 years depending on how often the scales
are used.
8.6 Storage and Transport Conditions
Keep all original packaging materials and components for eventual return of the scale to avoid
damage during transport; these are not covered by the warranty.
Disconnect all cables before transport and use the mains switch on the back of the scale to switch
it off to avoid damage.