Owners Guide
28
Turn ever ything o . Check to see that all cables are firmly seated and screwed
in. BNC cables from the GSR Amp must be connected to a positive input on the
PowerLab. Make sure the input is the same channel from which you expect to use
the front-end in the so ware. A er checking connections, restart the PowerLab
and application to see if this has fixed the problem.
You are using an early version of LabChart.
Upgrade to the latest version of the so ware. Contact your ADInstruments
representative for information.
The BNC or I
2
C cable is faulty.
Replace the cable and try again. Immediately label all cables proved faulty so
that you don’t use them again by accident.
The front-end is faulty.
This is the least likely event. If the front-end will not work properly a er the
previous measures, then try using it on another PowerLab. If the same problems
recur with a second PowerLab, the front-end may be faulty. Contact your
ADInstruments representative to arrange for repairs.
On starting up the so ware, an alert indicates that there is a problem with the front-end or
driver
The correct BP/GSR drivers are not installed on your computer.
Reinstall the so ware.
You are using an early version of LabChart.
Upgrade to the latest version of the so ware. Contact your ADInstruments
representative for information.
The BNC or I
2
C cable is faulty.
Replace the cable and try again. Immediately label all cables proved faulty so
that you don’t use them again by accident.
The front-end is faulty.
This is the least likely event. If the front-end will not work properly a er the
previous measures, then try using it on another PowerLab. If the same problems
recur with a second PowerLab, the front-end may be faulty. Contact your
ADInstruments representative to arrange for repairs.
Some so ware settings don’t resemble those in this guide
You are using an early version of the front-end driver, or of LabChart. Some changes
may have been made since then.
Upgrade to the latest version of the so ware. Contact your ADInstruments
representative for information.