FL20X Instrument Setup
Document # S200-FL20X-004
Rev 1.2 12/22/2014
Page 2 of 2
Aerospace Logic Inc.
Tel. 416-628-0725
www.aerospacelogic.com
4. To enter parameters and perform the calibration, proceed as follows:
a. Turn off the power to the instrument
b. Press and hold both top and bottom buttons and turn on the power to the
instrument.
c. When instructed, release both buttons.
d. In the setup menus, use the bottom button to navigate and the top button
to select an option.
e. Follow the onscreen instructions to enter the correct operating
parameters. NOTE: Depending on the installation orientation of the
instrument the display may initially be upside down. The first prompt
in the setup is to press the button at the top of the instrument. This
will orientate it correctly.
f. To perform the instrument calibration, follow the instructions provided on
the instrument screen.
5. Once all calibration has been completed the instrument will be enabled and be
ready for use.