74
2312A EN 20040909
Remarks for HF-emission and immunity
This device is intended for operation in the electromagnetic
environment given below. The user of the device should ensure that
it is used in such an environment.
This device was tested for a normal hospital environment as described above.
Nevertheless the HF-emission and immunity can be influenced by connected data
cables depending on length and the manner of installation.
Transmission
Measurements
Agreement
Electromagnetic Environment
Guidelines
High frequency transmissions in
accordance with CISPR 11
Group 1
The device uses high frequency energy
exclusively for its internal functions. For
this reason, its high frequency transmission
is very low and it is improbable that
neighboring electronic equipment will be
disrupted.
High frequency transmissions in
accordance with CISPR 11
Class B
The device is intended for use in all
buildings, including living areas and areas
directly connected to a public supply
network that also supplies buildings that
are used for domestic purposes.
Excess oscillations in accordance
with IEC 61000-3-2
Class A
Voltage fluctuations /
flickering in accordance with
IEC 61000-3-3
Fulfilled