Calibration Procedures 3
34980A Service Guide 65
Gain Adjustment Procedure
Adjustment for each function should be performed only in the order shown in the
performance verification table. The performance verification tables used for gain
adjustments start on page 57.
Review the “DMM Test Considerations" on page 55 and “Gain Adjustment
Considerations" on page 63 sections before beginning this test.
Configuration: DC functions — 6½ digits
AC functions — LF 3 HZ:SLOW
1 Configure each function and range shown in the gain verification tables (starting on
page 57).
2 Apply the input signal shown in the “Input” column of the appropriate verification
table.
.
3 Send the actual input value to the instrument using the CALibration:VALue
<value>
command.
4 Calibrate the instrument using the CALibration? command.
5 Perform the appropriate Gain Verification Test to check the calibration results.
6 Repeat steps 1 through 5 for each gain verification test point shown in the tables.
NOTE
Always complete tests in the same order as shown in the appropriate
verification table.
NOTE
Each range in the gain adjustment procedure takes less than 20 seconds
to complete.