Calibration Procedures 3
34980A Service Guide 67
Plug-in Modules
Plug-in Module Test Considerations
For optimum performance, all test procedures should comply with the following
recommendations:
Assure that the calibration ambient temperature is stable and
between 18 °C and 28 °C. Ideally the calibration should be
performed at 23 °C ± 1 °C.
Assure ambient relative humidity is less than 80%.
Install the plug-in module and allow a 1 hour warm-up period before verification or
adjustment.
Use shielded twisted pair PTFE insulated cables to reduce settling and noise errors.
Keep the input cables as short as possible.
Remove all user wiring and connections from the plug-in modules before verification
or adjustment.
Use 4-wire Ohms measurement techniques for checking relay contact resistance.
Check directly at the terminals where possible.