Calibration Procedures 3
34980A Service Guide 71
.
[1] Apply a measured “0” offset to this measurement.
Output Voltage Quick
Check
Error from
Nominal (90 day)
16 V Q ± 11 mV
12 V ± 9 mV
10 V ± 8 mV
8 V ± 7 mV
4 V ± 5 mV
0 V Q ± 3 mV
-4V ± 5 mV
-8 V ± 7 mV
-10 V ± 8 mV
-12 V ± 9 mV
-16 V Q ± 11 mV
Output Current Quick
Check
Error from
Nominal (90 day)
20 mA Q ± 23 A
15 mA ± 18.5 A
10 mA ± 14 A
5 mA ± 9.5 A
0 mA
[1]
Q ± 5 A
-5 mA ± 9.5 A
-10 mA ± 14 A
-15 mA ± 18.5 A
-20 mA Q ± 23 A