Chapter 4 Retrofitting Options
To Retrofit A5 DC Power Input Assembly (Option 002)
Assembly-Level Service Guide 4-5
4
Retrofitting Procedure
1 Slide A1 Motherboard Assembly into the new chassis (MP10) as shown in
Figure 4-1A.
Figure 4-1A. DC Power Input Assembly Retrofitting
2 Insert but do not tighten the three TORX 15 screws (H1) that attach A1
Motherboard Assembly to the new chassis (MP10) as shown in Figure 4-1A.
3 Insert and tighten the three BNC nuts (H3) onto the BNC connectors that
protrude from the rear of the chassis using the 14-mm deep-socket spin tight.
4 Insert and tighten the two hex screws (H4) and the lock washers (H5) onto the
GPIB connector that protrudes from the rear of the chassis using the 7-mm spin
tight.
H4
H1
H1
H1
H5
H3
A1 Motherboard
Assembly
H5
MP10