Chapter 1 Performance Tests
Test 1: Time Interval (Agilent 53131A/132A Only)
1-18 Assembly-Level Service Guide
1
Procedure
1 Connect the pulse generator signal to Channel 1 of the Counter as shown in
Figure 1-5.
Figure 1-5. Time Interval (Channel 1 TO 2) Test Setup
2 Set the Pulse Generator to the following:
PERIOD: 10 µs
WIDTH: 150 ns
High: 1.00 V Low : 0.00 V
Input Mode: Normal
3 On the Counter, Press Single/Stop key.
The Agilent 53131A Counter should display 0.1500 µs ±0.0040 µs.
The Agilent 53132A Counter should display 0.1500 µs ±0.0025 µs.
4 Record this value in the appropriate place in the Performance Test Record (Test
1, Line 1 for Agilent 53131A or Test 1, Line 2 for Agilent 53132A).
5 Disconnect the test setup.