EasyManua.ls Logo

Agilent Technologies 53131A - Page 43

Agilent Technologies 53131A
252 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1 Performance Tests
Test 2: Trigger Level (Agilent 53131A/132A Only)
Assembly-Level Service Guide 1-19
1
Test 2: Trigger Level (Agilent 53131A/132A Only)
This test verifies the Trigger Level accuracy of the Agilent 53131A and
Agilent 53132A Universal Counters.
Equipment
Agilent 3325B Synthesizer
Agilent 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2 Set CHANNEL 1 input conditions to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: 0.000V
50/1M, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
3 Set CHANNEL 2 input conditions to:
Trigger/Sensitivity, AUTO TRG: OFF
Trigger/Sensitivity, LEVEL: 0.000V
50/1M, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
深圳德标仪器 135-1095-0799

Related product manuals