Agilent 5100/5110 and 5800/5900 ICP-OES
Easy-fit Fully Demountable Torch User Guide
Introduction
The ICP-OES torch is a crucial component of the sample introduction system for the ICP-OES
spectrometers. It plays an important role in determining the sensitivity, precision and stability achieved
during analyses.
Figure 1: Fully demountable torch assembly
This document outlines the recommended procedures for assembly and disassembly, installation, cleaning
and maintenance of the fully demountable torch.
The fully demountable torch is composed of five main parts:
Figure 2 Exploded view
1. Locking nut
2. Injector
(quartz or inert alumina)
3. Torch base
4. Top seal
5. Quartz outer
tube-set
The fully demountable torch kit is supplied with all the components required for torch assembly. Some kits
include a spare quartz outer tube-set. The torch can be used in either RV or DV mode, when fitted with the
appropriate quartz outer tube-set.
The torch design allows either the injector and/or the quartz outer tube-set to be removed from the torch
base for easy cleaning or replacement.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5800 ICP-OES and is the answer not in the manual?

Agilent Technologies 5800 ICP-OES Specifications

General IconGeneral
BrandAgilent Technologies
Model5800 ICP-OES
CategoryFlashlight
LanguageEnglish

Related product manuals