52 Chapter2
Making Adjustments
5. Crystal and LC Bandwidth Filter
15.Remove the crystal shorts and press the following analyzer keys.
PEAK SEARCH
MKR FCTN, MK TRACK ON OFF (ON)
SPAN, 50, kHz
MKR FCTN, MK TRACK ON OFF (OFF)
BW, 3, kHz
PEAK SEARCH, MARKER DELTA
BW, 30, kHz
PEAK SEARCH
16.Verify that the MARKER frequency does not exceed 3 kHz.
If the signal shift is out of tolerance, repeat step 3 to step 16.
17.Press the following analyzer keys.