92 Chapter2
Making Adjustments
14. Comb Generator
14. Comb Generator
This adjustment applies to:
8592L spectrum analyzers
8593E spectrum analyzers
8596E spectrum analyzers
The output signal from the A3A1 comb generator assembly, with the
Step Recovery Diode Module (SRD) disconnected, is adjusted for
maximum peak-to-peak voltage. A3A1C5 FREQ is centered, and the
comb generator frequency is measured with a frequency counter. If the
measured frequency is not 100.000 MHz ±0.0004 MHz, A3A1L3 is
selected to bring the frequency within tolerance.
The comb generator signal is adjusted for maximum output power as
measured with a measuring receiver (used as a power meter). If the
amplitude is not +26.0 ±0.8 dBm, A3A1R6 is selected to bring the
amplitude within tolerance.
A3A1C5 FREQ is adjusted for a comb generator frequency of
100.000000 MHz ±0.000010 MHz (tolerance of ±10 Hz).
Equipment Required
Oscilloscope
Frequency counter
Measuring receiver (used as a power meter)
Power sensor, 300 MHz
Attenuator, 20 dB
Cable, SMA, 90 cm (36 in.)
Cable, BMC, 120 cm (48 in.)
Adapter, Type N (m) to SMA (f)
Adapter, Type N (f) to Type N (f)
Adapter, Type N (m) to BNC (f)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8591E and is the answer not in the manual?

Agilent Technologies 8591E Specifications

General IconGeneral
BrandAgilent Technologies
Model8591E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals