Chapter 2 91
Making Adjustments
13. Third Converter and Second IF Bandpass
For example, if the Corrected Insertion Loss is 10.0 dBm, and the
measuring receiver reading is 30 dBm, then 30 dBm (10.0)
dBm = 20 dBm.
CAL OUT Power____________________dBm
32.Adjust A9R19 CAL OUT ADJ accordingly if the CAL OUT amplitude
is not 20 dBm ±0.4 dB as calculated in step 26.