90 Chapter2
Making Adjustments
13. Third Converter and Second IF Bandpass
28.Subtract the Mismatch Error (step 22) from the Uncorrected
Insertion Loss (step 22). This is the corrected insertion loss.
Corrected Insertion Loss____________________dB
For example, if the Mismatch Error is +0.3 dB and the uncorrected
Insertion Loss is 10.2 dB, subtracting the mismatch error to the
insertion loss gives a corrected reading of 10.5 dB.
300 MHz Calibrator Amplitude Adjustment
29.Connect the equipment as shown in Figure 2-24. The analyzer
should be positioned so that the setup of the adapters, LPF and
attenuator do not bind. It may be necessary to support the center of
gravity of the devices.
Figure 2-24 300 MHz Calibrator Amplitude Accuracy Test Setup
30.On the power meter, press the dBm mode key. Record the power
meter reading in dBm.
Power Meter Reading____________________dBm
31.Subtract the Corrected Insertion Loss (step 23) from the power
meter reading (step 25) and record as the CAL OUT power. The CAL
OUT should be 20 dBm ±0.4 dB.
CAL OUT Power = Power Meter Reading Corrected Insert Loss

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8591E and is the answer not in the manual?

Agilent Technologies 8591E Specifications

General IconGeneral
BrandAgilent Technologies
Model8591E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals