Chapter 2 97
Making Adjustments
15. Frequency Response of the 8590L, 8591C, and 8591E
Procedure for System Characterization (75
input)
1. Zero and calibrate the measuring receiver and 1 MHz to 1.8 GHz
power sensor as described in the measuring receiver operation
manual.
2. Zero and calibrate the power meter and the 75 power sensor as
described in the power meter operation manual.
3. Press INSTRUMENT PRESET on the synthesized sweeper. Set the
synthesized sweeper controls as follows:
CW .................................................................. 41MHz
FREQ STEP ................................................... 37MHz
POWER LEVEL ............................................... 5dBm
4. Connect the equipment as shown in Figure 2-27.
Figure 2-27 System Characterization Test Setup for 75 inputs
CAUTION Use only 75 cables, connectors, or adapters on instruments equipped
with 75 inputs or damage to the input connectors will occur.
5. Adjust the synthesized sweeper POWER LEVEL for a 0 dBm
reading on the measuring receiver.
6. Record the power meter reading in Column 4 of Table 2-6, taking
into account the cal factors of both power sensors.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8594E and is the answer not in the manual?

Agilent Technologies 8594E Specifications

General IconGeneral
BrandAgilent Technologies
Model8594E
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals