12
Test Port Output Characteristics
Frequency characteristics
Range 300 kHz to 3 GHz
Option 006 30 kHz to 6 GHz
Resolution 1 Hz
Stability typically ±7.5 ppm 0° to 55° C
typically ±3 ppm/year
Option 1D5 typically ±0.05 ppm 0° to 55° C
typically ±0.5 ppm/year
Accuracy ±10 ppm at 25° ± C
Power range -5 to +20 dBm
Option 006 -5 to +18 dBm
Resolution 0.01 dB
Level accuracy
1,2
±1.0 dB
Level linearity
1,2,5
±0.25 dB, -5 to +15 dBm
±0.5 dB, +15 to +20 dBm
Impedance 50 nominal
300 kHz to 3 GHz
2
>16 dB RL (<1.38 SWR)
300 kHz to 6 GHz >14 dB RL (<1.50 SWR)
Spectral purity
2nd harmonic
3
<-40 dBc at +10 dBm
6
<-50 dBc at 0 dBm
6
Option 002 <-25 dBc at max power
3rd harmonic
4
<-40 dBc at +10 dBm
6
<-50 dBc at 0 dBm
6
Option 002 <-25 dBc at max power
Nonharmonic spurious
6
Mixer related <-30 dBc at max power
<-55 dBc at -10 dBm
Test Port Input Characteristics Option 011
Frequency range 300 kHz to 3 GHz
Option 006 30 kHz to 6 GHz
Average noise level
2
50 kHz to 3 GHz <-90 dBm (3 kHz BW)
50 kHz to 3 GHz <-110 dBm (10 Hz BW)
50 kHz to 3 GHz <-120 dBm (10 Hz BW) typical
3 to 6 GHz <-85 dBm (3 kHz BW)
3 to 6 GHz <-105 dBm (10 Hz BW)
3 to 6 GHz <-115 dBm (10 Hz BW) typical
Maximum input level 0 dBm
Damage level 20 dBm or 25 VDC
Impedance 50 ohms nominal
300 kHz to 2 MHz 20 dB RL
2 MHz to 1.3 GHz 24 dB RL
1.3 GHz to 3 GHz 19 dB RL
Option 006
3 GHz to 6 GHz 15 dB RL
6
Frequency response
300 kHz to 3 GHz ±1.0 dB
3 GHz to 6 GHz ±2.0 dB
Harmonics (Option 002)
2nd harmonic
3
<-15 dBc at 0 dBm
<-30 dBc at -10 dBm
6
<-45 dBc at -30 dBm
6
3rd harmonic
4
<-30 dBc at 0 dBm
<-50 dBc at -10 dBm
6
<-50 dBc at -30 dBm
6
Harmonic measurement accuracy
16 MHz to 3 GHz ±1.5 dB
3 GHz to 6 GHz Option 006 ±3 dB
Harmonic measurement dynamic range
-40 dBc
6
(output = -10 dBm, input <-15 dBm)
Agilent 8753ES Option 011 Specifications
1. +10 dBm output power for Agilent 8753ES Option 011.
2. Typical below 300 kHz.
3. 16 MHz to 3 GHz.
4. 16 MHz to 2 GHz.
5. For Agilent 8753ES Option 011 and Option 006, linearity is specified for the
ranges of -5 to +13 dBm and +13 to +18 dBm.