Chapter 2 2-173
System Verification and Performance Tests
Agilent 8753ET Performance Test Records
Agilent Technologies Company
Model 8753ET
Serial Number
Report Number
10. Uncorrected Port Performance
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
Test Description Frequency Range Spec.
(dB)
Measured Value
(dB)
Meas. Uncert.
(dB)
Forward Direction
Directivity 300 kHz to 1.3 GHz 30 ± 0.9
Directivity 1.3 GHz to 3 GHz 24 ± 0.6
Directivity 3 GHz to 6 GHz 19 ± 0.8
Forward Direction
(Standard)
Source Match 300 kHz to 1.3 GHz 25 ± 1.3
Source Match 1.3 GHz to 3 GHz 20 ± 1.5
Source Match 3 GHz to 6 GHz 14 ± 1.3
Forward Direction
(Option 004)
Source Match 300 kHz to 1.3 GHz 23 ± 1.0
Source Match 1.3 GHz to 3 GHz 18 ± 1.2
Source Match 3 GHz to 6 GHz 14 ± 1.3