Chapter 2 2-175
System Verification and Performance Tests
Agilent 8753ET Performance Test Records
Agilent Technologies Company
Model 8753ET
Serial Number
Report Number
11. System Trace Noise
Test Port IF BW Phase/Magnitude Measured Value Specification
Test Frequency: 3 GHz
Reflection 3 kHz Magnitude <0.006 dB rms
Reflection 3 kHz Phase <0.038 °rms
Reflection 10 Hz Phase <0.006 °rms
Reflection 10 Hz Magnitude <0.001 dB rms
Transmission 10 Hz Magnitude <0.001 dB rms
Transmission 10 Hz Phase <0.006 °rms
Transmission 3 kHz Phase <0.038 °rms
Transmission 3 kHz Magnitude <0.006 dB rms
Test Frequency: 6 GHz
Note: If your analyzer does not have Option 006, write “N/A” in all entries above 3 GHz.
Reflection 3 kHz Magnitude <0.010 dB rms
Reflection 3 kHz Phase <0.070 °rms
Reflection 10 Hz Phase <0.012 °rms
Reflection 10 Hz Magnitude <0.002 dB rms
Transmission 10 Hz Magnitude <0.002 dB rms
Transmission 10 Hz Phase <0.012 °rms
Transmission 3 kHz Phase <0.070 °rms
Transmission 3 kHz Magnitude <0.010 dB rms

Table of Contents

Other manuals for Agilent Technologies 8753ES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8753ES and is the answer not in the manual?

Agilent Technologies 8753ES Specifications

General IconGeneral
BrandAgilent Technologies
Model8753ES
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals