Chapter 5 Error Messages
Execution Errors
130
-221 Settings conflict
Indicates that a legal program data element was parsed but could not be executed due
to the current device state.
-222 Data out of range
A numeric parameter value is outside the valid range for the command.
Example: TRIG:DEL -3
-223 Too much data
A character string was received but could not be executed because the string length
was more than 40 characters. This error can be generated by the
CALibration:STRing command.
-224 Illegal parameter value
A discrete parameter was received which was not a valid choice for the command.
You may have used an invalid parameter choice.
Example: DISP:STAT XYZ (XYZ is not a valid choice).
-330 Self-test failed
The power supply’s complete self-test failed from the remote interface (*TST?
command). In addition to this error, more specific self-test errors are also reported.
See also ‘‘Self-Test Errors’’, starting on page 133.
-350 Queue overflow
The error queue is full because more than 20 errors have occurred. No additional errors
are stored until you remove errors from the queue. The error queue is cleared when
power has been off, or after a *CLS (clear status) command has been executed.
-410 Query INTERRUPTED
A command was received which sends data to the output buffer, but the output buffer
contained data from a previous command (the previous data is not overwritten). The
output buffer is cleared when power has been off, or after a *RST (reset) command
has been executed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3646A and is the answer not in the manual?

Agilent Technologies E3646A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3646A
CategoryPower Supply
LanguageEnglish

Related product manuals