EasyManuals Logo

Agilent Technologies E4411B User Manual

Agilent Technologies E4411B
321 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #208 background imageLoading...
Page #208 background image
6-16 Chapter6
Front-Panel Key Reference
BW/Avg
EMI Res BW Accesses the EMI Res BW menu keys and allows you to choose between
bandwidths of
120 kHz, 9 kHz and 200 Hz. The 200 Hz bandwidth is only
available if Option 1DR (narrow resolution bandwidth) is installed and
the span is less than 5 MHz. (Option 1DR is available with Agilent
ESA-E Series only (E4401B, E4402B, E4404B, E4405B and E4407B.)
This function is set to
None when the resolution bandwidth is set to any
other value using the
Resolution BW key.

Table of Contents

Other manuals for Agilent Technologies E4411B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4411B and is the answer not in the manual?

Agilent Technologies E4411B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4411B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals