Hardkey and Softkey Reference ESG Family Signal Generators
Utility
4-130 User’s Guide
RCDMA
Press this softkey to set the current catalog type to RCDMA and display
the catalog of RCDMA files.
Softkey Location:
Utility > Memory Catalog > Catalog Type >
ARB Catalog Types > RCDMA
Recall Ref Osc Setting
This softkey is provided for adjusting the internal reference oscillator.
This adjustment is for use by service personnel only. Refer to the
calibration guide for more information.
Softkey Location:
Utility > Instrument Adjustments >
Reference Oscillator Adjustment > Recall Ref Osc Setting
Receive Pace None Xon
This softkey in the RS-232 Pace menu determines whether any pacing
is acknowledged by the RS-232 receive channel. If Xon pacing is used,
the system will send an Xoff character to suspend the transmitting of
data when its buffer is nearly full. When the system is ready to receive
further input, an Xon character is sent to the transmitting device. Xon
pacing is not recommended if binary data will be queried from the
signal generator via RS-232. Toggle
ReceivePaceNoneXon to None to turn
off Xon pacing.
Pace is a persistent state; it is not affected by an instrument preset or
by a power cycle.
Softkey Location:
Utility > GPIB/RS-232 > RS-232 Pace >
Receive Pace None Xon
Reference Oscillator Adjustment
Pressing this softkey reveals a menu of softkeys for adjusting the
internal reference oscillator. This adjustment is for use by service
personnel only.
Softkey Location:
Utility > Instrument Adjustments >
Reference Oscillator Adjustment
Ref Osc Coarse
This softkey is provided for adjusting the internal reference oscillator.
This adjustment is for use by service personnel only. Refer to the
calibration guide for more information.
Softkey Location:
Utility > Instrument Adjustments >
Reference Oscillator Adjustment > Ref Osc Coarse

Other manuals for Agilent Technologies E4422B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4422B and is the answer not in the manual?

Agilent Technologies E4422B Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4422B
CategoryPortable Generator
LanguageEnglish

Related product manuals