Installation Note E4440-90597 3
Option BAB Hardware Retrofit Kit
Option BAB Hardware Retrofit Kit
Introduction
The Option BAB retrofit kit provides all the parts and instructions necessary to install
the APC 3.5 input connector into the Agilent E4440A spectrum analyzer. The Option
BAB is licensed for one instrument model number/serial number combination. The kit
will only function on the instrument it was ordered for.
NOTE 1. The installation of this kit requires that some re-adjustment and
performance testing be performed in order to assure the new option is
functioning properly. This installation note includes a list of required
adjustments and performance tests. Completing the list of required
performance tests does not guarantee the instrument meets all
specifications.
NOTE 2. The instrument end user must determine whether they need a full
instrument calibration following the installation of the kit. If full
calibration is required, arrangements regarding the level of calibration
must be made between the end user and the calibration provider.
Products Affected: PSA E4440A
Serial Numbers: US00000000/US99999999
MY00000000/MY99999999
To Be Performed By: (X) Agilent Service Center
(X) Personnel Qualified by Agilent
( ) Customer
Estimated Installation Time:
Estimated Adjustment and Verification Time for
Hardware Kit following Installation:
45 Minutes
3 Hours (see note 1)
Additional Recommended Task Agilent recommends that a full calibration be
performed to verify instrument specifications.
(see note 2)