EasyManuals Logo
Home>Agilent Technologies>Inverter>E8267D

Agilent Technologies E8267D User Manual

Agilent Technologies E8267D
366 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #162 background imageLoading...
Page #162 background image
152 Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide
Programming the Status Register System
Accessing Status Register Information
*ESR? (event status register) queries and clears the Standard Event Status Register which is part
of the Standard Event Status Group.
*OPC, *OPC? (operation complete) sets bit #0 in the Standard Event Status Register to 1 when all
commands have completed. The query stops any new commands from being processed until the
current processing is complete, then returns a 1.
*PSC, *PSC? (power- on state clear) sets the power- on state so that it clears the Service Request
Enable Register, the Standard Event Status Enable Register, and device- specific event enable
registers at power on. The query returns the flag setting from the *PSC command.
*SRE, *SRE? (service request enable) sets and queries the value of the Service Request Enable
Register.
*STB? (status byte) queries the value of the status byte register without erasing its contents.
:STATus:PRESet presets all transition filters, non- IEEE 488.2 enable registers, and error/event
queue enable registers. (Refer to Table 4- 2.)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E8267D and is the answer not in the manual?

Agilent Technologies E8267D Specifications

General IconGeneral
BrandAgilent Technologies
ModelE8267D
CategoryInverter
LanguageEnglish

Related product manuals