EasyManua.ls Logo

Agilent Technologies EasyEXPERT

Agilent Technologies EasyEXPERT
610 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 PwrDevice
10.8 Vth gmMax[3] PwrDevice: Linear region Vth (A.01.20)
[Supported Analyzer]
B1500A
[Description]
Extracts the threshold voltage (Vth) by using the extrapolation method for the linear region of the drain current
vs gate voltage characteristics.
[Device Under Test]
Power MOSFET, 3 terminals
[Device Parameters]
Polarity: Nch (SMUs force the specified value) or Pch (SMUs force the negative specified value).
Lg: Gate length
Wg: Gate width
Temp: Temperature
IdMax: Drain current compliance
[Test Parameters]
IntegTime: Integration time
Gate: SMU connected to Gate terminal, primary sweep voltage output
VgStart: Sweep start voltage for Gate terminal
VgStop: Sweep stop voltage for Gate terminal
VgStep: Sweep step voltage for Gate terminal
Drain: SMU connected to Drain terminal, constant voltage output
Vd: Drain voltage, ideally at around 100mV
Source: SMU connected to Source terminal, constant voltage output
BaseOffsetV: Base offset voltage
Base offset voltage is added to the specified voltage. For example, the gate start voltage will be
VgStart+BaseOffsetV.
[Extended Test Parameters]
Vs: Source voltage
IgLimit: Gate current compliance
HoldTime: Hold time
DelayTime: Delay time
gm_Min: Minimum gm value for graph scale
gm_Max: Maximum gm value for graph scale
DrainMinRng: Minimum range for drain current measurement
[Measurement Parameters]
Drain current Idrain
[User Function]
gm=diff(Idrain,Vgate)
[Analysis Function]
gmMax=max(gm)
Von=@L1X (X intercept of Line1)
Vth=Von-Vd/2
Vth is given by the following formula.
Agilent EasyEXPERT Application Library Reference, Edition 8
10-15

Related product manuals