10 PwrDevice
Vth=Vg(gmMax)-Id(gmMax)/gmMax
Vd/2 is for compensation of the secondary term of Vd in the theoretical formula.
[Auto Analysis]
Line1: Tangent line for Y1 at gm=gmMax
[X-Y Plot]
X axis: Gate voltage Vgate (LINEAR)
Y1 axis: Drain current Idrain (LINEAR)
Y2 axis: Transconductance gm (LINEAR)
Y3 axis: Drain current Idrain (LOG)
[List Display]
Gate voltage Vgate
Source voltage Vsource
Drain voltage Vdrain
Drain current Idrain
Transconductance gm
[Parameters Display Area]
Threshold voltage Vth
Maximum transconductance value gmMax
Agilent EasyEXPERT Application Library Reference, Edition 8